The Headwall Photonics Hyperspec™ family of NIR and VNIR integrated hyperspectral imaging sensors provide the foundation for utilizing hyperspectral imaging to achieve superior spectral sensing and chemical imaging results for a range of mission-critical applications ranging from process monitoring to moving webs of product across conveyor lines to non-invasive medical imaging.
Headwall Photonics’ Hyperspec™ imaging spectrometer platform is built on a totally reflective concentric, f/2 optical design. These products are based on Headwall’s patented aberration-corrected imaging systems which feature the company’s “original”, high efficiency holographic gratings.
In order to minimize stray light and aberrations, the use of transmissive optical components are not used within the imaging spectrograph. This platform is further enhanced by a telecentric optical input design which enables superior spatial imaging.
Select the Hyperspec™ model which best satisfies spectral range, desired illumination distance, and spatial resolution. Then select the entrance slit width which best provides your desired spectral resolution; the standard slit width is 12.5 microns wide. By choosing a wider slit width, spectral resolution will be approximately broadened by a factor of the increased dimension (example: 25 µm = 2x). The objective lenses on Hyperspec imaging sensors are field interchangeable, providing the ability to vary the FOV and IFOV within the application scene.
|
Hyperspec-NIR™ X Series Performance Guide |
|
|
Spectral / Optical |
XF Series |
|
λ Coverage |
900 - 1700 nm |
|
Spectral Bandpass¹ |
2 nm/pixel |
|
Spectral Dispersion |
4.8 nm/pixel |
|
Nyquist Resolution |
9.6 nm |
|
FWHM of Slit Image |
4 nm (spectral resolution based on 25µm slit) |
|
Electronics |
|
|
Frame Rate |
60 fps (USB2), 100 fps (LVDS) |
|
Cooling |
Forced convection |
|
Exposure Time |
1 µsec – 5 minutes |
|
Digital Resolution |
12bit @ 60Hz and 100Hz, 14bit @ 350Hz |
|
Data Interface |
USB 2.0, Camera Link, LVDS |
|
Dynamic Range |
69dB low gain, 60dB high gain |
|
Window of Interest |
Smallest window 8x32 pixels (30x30 microns) |
|
Trigger, Auxiliary Signals |
TTL level triggers |
|
Power Consumption |
< 4 W, cooler: 30 W max |
|
Power Supply |
12V – 5A |
|
CE Certified |
Yes |
|
Spatial Imaging Performance Guide |
|||||||||
|
Model |
Scene¹ FOV focused at: |
Scene¹ IFOV focused at: |
Scene Res.² focused at: |
||||||
|
|
∞ (º) |
0.75 m (mm) |
3.0 m (mm) |
∞ (mrad) |
0.75 m (mm) |
3.0 m (mm) |
∞ (mrad) |
0.75 m (mm) |
3.0 m (mm) |
|
HS-NIR-XF-25 |
21.74 |
278 |
1142 |
1.20 |
0.87 |
3.57 |
2.40 |
1.74 |
7.14 |
|
HS-NIR-XF-35 |
15.62 |
196 |
813 |
0.86 |
0.61 |
2.54 |
1.43 |
1.23 |
5.98 |
|
|
|
3.0 m (mm) |
5.0 m (mm) |
|
3.0 m (mm) |
5.0 m (mm) |
|
3.0 m (mm) |
5.0 m (mm) |
|
HS-NIR-XF-50 |
10.97 |
566 |
950 |
0.60 |
1.77 |
2.97 |
1.20 |
3.54 |
5.94 |
|
HS-NIR-XF-75 |
7.32 |
374 |
630 |
0.40 |
1.17 |
1.97 |
0.80 |
2.34 |
3.94 |
|
HS-NIR-XF-100 |
5.50 |
278 |
470 |
0.30 |
0.87 |
1.47 |
0.60 |
1.74 |
2.94 |
|
Lens Number |
f/Stop Range |
Min Obj Dist (w/o ext. tube) |
|
HS-NIR-Lens-25 |
2.0 - 16 |
300 mm |
|
HS-NIR-Lens-35 |
2.0 - 16 |
700 mm |
|
HS-NIR-Lens-50 |
2.0 - 16 |
1.5 m |
|
HS-NIR-Lens-75 |
2.0 - 16 |
2.5 m |
|
HS-NIR-Lens-100 |
2.0 - 16 |
3 m |
HS-NIR objective lenses are designed for 900–1700 nm precision imaging. Premium optical performance is achieved through a sophisticated design using ultra low dispersion glass and special broadband anti-reflection coatings. These optimized lenses further enhance the signal throughput and low stray light performance of the Hyperspec™ imaging platform.
The HS-NIR-XF series Hyperspectral Imaging Camera interfaces to a PC via a standard USB 2.0, Cameralink or a parallel LVDS bus.
The HS-NIR-XF series Hyperspectral Imaging Camera interfaces to a PC via a standard USB 2.0, Cameralink or a parallel LVDS bus.